Select Page

UEP 15

The UEP 15 is a plug in power supply for use in UEN 04 bin or for powering every non-powered extant NIM bin. The UEP 15 features all 6 voltages, stabilized by adjustable monolithic voltage regulators. Outputs are ported using a rear mounted NIM connector with a standard NIM pin out.

An integrated DC blower provides perfect cooling of the unit to avoid overheating.  Voltage regulators and transformer are protected against over temperature.  Unique design of Over/Under voltage discriminator, maintains proper output regulation within +/- 3% of nominal voltages.  Should the voltage fall out of this range the mains relay shuts down the power supply with a, “Bad Status”, (red LED indicator on front panel).

Optional CAN-bus interface for DC voltage monitoring and remote on/off via rear 9 pin sub-D connector.

Features

UEP 15 power supply:

  • Front side plug-in slot power supply 2.5 NIM slots wide, 150W power output
  • 6-fold DC outputs (+/-6V, +/-12V, +/-24V), stabilized by adjustable monolithic voltage regulators switch-off protection for over / under voltage and over temperature failures
  • Power distribution rear side via standard NIM connector
  • Front panel with mains switch and control LED´s for status and failure
  • Optional with CAN-bus interface for remote monitoring and control
  • AC input 230V/50 Hz or 110V/60Hz
  • Dimensions: width 2 ½  NIM, height 183 mm, depth 249 mm, weight: 7.5 kg
  • CE Conformity

Ordering Information

Standard configurations (others possible by request).

Type+6V/-6V+12V/-12V+24V/-24V115VACPower
UEP15_x5A/5A3A/3A1.5A/1.5A0.25A150W

x = defines the AC input voltage, factory default is 220V AC (without index)
x = B: 110V AC
x = J: 100V AC
x = E: 240V AC

Technical Information

NIM Compact Slot Power Supply UEP 15

Plug in power supply for use in UEN 04 bins or for powering every non-powered extant NIM bin. UEP 15 features all 6 voltages, stabilized by adjustable monolithic voltage regulators, on the rear NIM connector with standard NIM pin out. Ground and 6V outputs are spliced-wired to two pins for minimized voltage drop.

The integrated blower cools heat sink, components and the toroid transformer. Voltage regulators and transformer are protected against over temperature. Over- under voltage detector (limits ca. +/-3% of nominal voltages) activates mains cutting relay (red signal "bad status" at front panel).

AC Mains voltages: 100, 115, 225, 240VAC by use of different internal transformer taps.

Power supply DC-Outputs: +6V -6V +12V -12V +24
UEP15 5A 5A 3A 3A 1,5A
UEP15 as 5A 5A 3A 3A 3A

 

Power supply DC-Outputs: -24V 115VAC max. power regulation application
UEP15 1,5A 0.1A 150W linear NIM
UEP15 as 0,5A 0.1A 150W linear NIM

 

UEP15  
Input voltage, 47-63Hz 100V, 110V, 215V, 220V, 240V all (+/-10%), 5 internal taps
Soft start yes
Output: Noise and Ripple: Full load +/-6 / 12 / 24V <3mVpp / <5mVpp / <10mVpp
Regulation static: Change of output voltage versus load change 10-100% <0,2% for 6V, 0,1% for 12/24V
Regulation static: change of output voltage versus line change  +/-10% 0,02%
Recovery time versus load change 10-100% <0,15ms
Output impedance:  Static / Dynamic(at 100kHz) 1,5mOhm / 0,5 Ohm
Temperature Error <0,02%/K
Thermal Protection (No. of thermal switches) (1x) and Regulators internally
Output Current Characteristics (Ishort <3A in foldback regulators), reverse bias diodes! Constant current and trip off
Dual tracking for complementary outputs no
Calibration ranges Voltage / Currents +/-5% / fix limits
Sense compensation  ranges, all DC voltages  0V
Status Control for all voltages (Over- Under-Voltage Comparator, defaults +/-0,3%) Bad Status red LED, unit trips off
Overvoltage Protection, trip off thresholds (defaults) +/-0,3% of nom voltages 
Derating >40°C with 2% up to 60°C max.

Optional CAN-bus interface for DC voltage monitoring and remote on/off via rear 9 pin sub D connector